Publication Date: 2/1/75
    Pages: 6
    Date Entered: 2/23/84
    Title: DESIGN CONSIDERATIONS - SYSTEMS FOR MEASURING THE MASS OF LIQUIDS
    February 1975
    U.S. NUCLEAR REGULATORY COMMISSION
    REGULATORY GUIDE
    REGULATORY GUIDE 5.48
    DESIGN CONSIDERATIONS
    SYSTEMS FOR MEASURING THE MASS OF LIQUIDS
A. INTRODUCTION
    Section 70.58, "Fundamental Nuclear Material Controls," of 10 CFR
    Part 70, "Special Nuclear Material," requires that the quantity of
    special nuclear material (SNM) on inventory be known on the basis of
    measurement at the beginning and end of each material balance interval.
    Section 70.51 of 10 CFR Part 70 requires that all SNM added to and
    removed from processing during that interval be measured. Section 70.58
    also requires that all random and systematic errors associated with such
    measurements be controlled so that the limit of error of material
    unaccounted for (LEMUF) at the end of any material balance interval does
    not exceed the amount specified in the regulation.
    This regulatory guide pertains to design considerations for
    methods of measuring the mass of liquid contained in a vessel and
    identifies those design considerations which the NRC staff considers to
    be adequate for minimizing the error associated with that measurement.
    Equipment and procedures for obtaining liquid samples are the subject of
    another regulatory guide, which is currently in preparation.
B. DISCUSSION
    Measurement of the SNM content of a vessel containing a
    homogeneous liquid solution consists of three principal operations: (1)
    measurement of liquid mass (either by weighing directly or by measuring
    the volume, temperature, and specific gravity); (2) obtaining a
    representative sample of the vessel contents; and (3) assay of the
    sample to determine the SNM concentration in terms of unit weight or
    unit volume.
    This guide deals with measurement activities that are affected by
    vessel design. Proper attention to design makes possible more accurate
    measurement of the solution bulk. It also improves the ease of
    achieving accurate bulk measurements and obtaining representative
    samples. The determination of chemical and isotopic concentrations in
    samples by traditional analytical techniques is independent of vessel
    design, and chemical and isotopic assay are not discussed in this guide.
1. Bulk Measurement of Liquid
    a. Measurement of Weight
    Determination of solution bulk by weighing is advantageous because
    this makes it unnecessary to measure the temperature and specific
    gravity of the solution, eliminating random and systematic errors
    inherent in measuring temperature and specific gravity and avoiding the
    possibility of mistakes in making temperature corrections. Another
    advantage of determining solution bulk by weight is that there is no
    requirement for dimensional uniformity within the vessel to achieve
    calibration linearity.
    A disadvantage of weighing is that rigid piping connected to the
    vessel can produce mechanical stresses due to deflection and thermal
    expansion. These stresses can affect the indicated weight. Weighing
    systems of the null-balance type, which restore the vessel to a
    reference position, can minimize the effect of stress caused by
    deflection of the vessel with increasing weight. However, such systems
    do not compensate for thermal expansion. Where disconnects are
    permissible and the number of connections is small enough to make
    disconnection feasible, the vessel may be disconnected from the piping
    during a weight determination to eliminate this effect. Otherwise,
    temperatures are held constant or an additional correction is included.
    In vessels equipped with heating-cooling jackets, errors in
    solution weight measurement can occur if the weight of fluid in the
    jacket varies. This is more likely to be a problem when steam rather
    than a liquid heat-transfer medium is used for heating. If a liquid
    heat-transfer medium is used, the problem may be less severe in a vessel
    with internal coils than in a vessel with an external jacket.
    Weighing a vessel containing SNM can be done by mounting the
    vessel on an industrial platform scale or by placing hydraulic load
    cells under the vessel supports. Industrial weighbeam balances and
    pendulum balances have capacities of tens of thousands of kilograms and
    accuracies of 0.1%.(1,2) Under carefully controlled conditions, an
    accuracy better than 0.1% can be achieved; 14,000-kg cylinders of
    uranium hexafluoride are routinely weighed with an accuracy of 0.01%.
    Pendulum balances indicate weight without requiring the manipulation of
    standard weights and thus are more readily adapted for remote reading
    and automatic printout.
    Industrial use of weighbeam or pendulum balances has been largely
    limited to applications where the vessel is accessible for maintenance
    and adjustment. Where radiation levels preclude such access, direct
    weighing can be accomplished by supporting the vessel on hydraulic load
    cells or by applying electric resistance-wire strain gages to the vessel
    supports.
    Hydraulic load cells have a capacity of about 25,000 kg.(2) An
    accuracy of 0.1% can be achieved.(3) The load on all the supports can
    be easily combined and can be read at any convenient remote location.
    The readings are sensitive to temperature, but substantial temperature
    correction is possible.
    The feasibility of using hydraulic load cells to measure the
    weight of process feed was studied at the Idaho Chemical Processing
    Plant.(3) Because the installation was on an existing vessel that was
    not adapted to the purpose, the tests were inconclusive. However,
    valuable insight was gained into desirable design features for such an
    installation.
    Electric strain gages are highly sensitive and have a capacity of
    200,000 kg. Strain-gage readings are influenced by temperature, but
    this effect can be readily compensated for. Their accuracy approaches
    that for hydraulic load cells, but they are expensive and relatively
    fragile; hence, they are not normally used for the weighing of vessels
    containing SNM.
    b. Measurement of Volume, Temperature, and Specific Gravity
    Bulk measurement of volume is usually accomplished indirectly:
    the liquid level is measured directly; then a prior calibration of
    liquid level vs. volume is applied. Removals or additions to process
    can be accurately measured by filling a precisely calibrated tank volume
    to overflowing and then transferring the known increment of volume to
    the process. Alternatively, liquid level can be measured by any of
    several devices: purged dip tubes, sight glass, or other types of level
    sensors.
    Pairs of pneumatic dip tubes are used in high radiation fields
    because of their simplicity. The opening of one of the tubes is
    submerged, and the opening of the other is in the vapor space above the
    liquid. The pressure difference between the two tubes while air is
    passed through the tubes at the same rate is proportional to the product
    of the height of liquid above the opening of the submerged tube and the
    density of the liquid. This pressure difference is sometimes called the
    "weight factor." For vessels having a uniform cross section, the weight
    factor is directly related to the mass of liquid above the submerged
    tube opening.
    Specific gravity can be measured by using a pair of dip tubes, the
    ends of which are submerged to different but known depths. However, the
    results are usually less reliable for accountability purposes than can
    be obtained by measuring the specific gravity of a representative sample
    withdrawn for chemical and isotopic analysis. The liquid level in a
    vessel is the quotient of the weight factor, expressed in suitable
    units, divided by the specific gravity of the liquid. When properly
    designed and installed,(4) a pair of purged dip tubes can measure liquid
    level with a precision of 0.6% of full scale. Accuracies of 0.1% to 0.2%
    can be achieved in a calibrated tank under carefully controlled
    conditions.
    A sight glass gives a positive reading of liquid level, but is
    subject to breakage (which could result in the spread of contamination).
    Sensing of liquid level can be done by other means, such as
    hydrometer-type floats or ultrasonic ranging. The resultant accuracy is
    generally not adequate for accountability measurements, however.
    In the design of vessels for volume measurement, the systematic
    error associated with the existence of a liquid heel below the lowest
    limit of the measuring device can be a serious problem. Heel
    uncertainty is not important for throughput measurements, where
    differences in liquid level are measured and the heel can be accounted
    for in the measurements. For measurements of absolute quantities during
    an inventory, however, the error introduced by the heel volume can
    constitute much of the measurement error. The complexity of volumetric
    calibration is greater when the vessel cross section varies with height,
    as may result if there are components (impellers, baffles, heat-cooling
    coils, etc.) within the vessel. On the other hand, effective mixing to
    obtain homogeneity generally requires some obstructions such as spargers
    or impellers and baffles. Hence, the designer seeks the trade-off
    features that are the most desirable for each application.
    Other drawbacks to determining bulk contents by measuring volume
    are the complications introduced by temperature changes and the errors
    contributed in making the necessary temperature corrections. Most
    vessels expand at elevated temperatures, although slab tanks may
    decrease in volume with increasing temperature if the sides are
    constrained to buckle inward.
    Not only does the vessel volume change with temperature, but the
    contained solution also expands when heated. Therefore, each
    liquid-level measurement has associated with it the temperature at which
    the measurement is made. Furthermore, laboratory determinations of
    specific gravity and SNM concentration are frequently made at
    temperatures other than those at which the bulk volume was measured and
    at which aliquots were taken. Therefore, the temperatures at which
    aliquots are taken should be recorded. Unless the uncertainties
    associated with temperature measurements are held within tolerable
    bounds and all temperature corrections are properly made, the accuracy
    of the bulk volume measurement will be poor.
2. Vessel Design Features Pertainent to Obtaining a Representative
    Sample(*) a. Mixing of Vessel Contents
    To ensure a liquid solution having uniform composition, vessel
    contents can be mixed by gas spargers, by mechanical impellers, or by
    external recirculation (pumping the solution out and back in).
    Gas spargers are relatively ineffective, lengthy sparging being
    required to bring about uniformity of the tank contents. They also
    increase the load on the offgas filters. However, they are inexpensive
    and require little or no maintenance. Spargers are more effective in
    tall, narrow tanks than in short, wide ones. The optimum gas flow rate
    is about 0.75 m(3)/sec (1.3 cfm) per square foot of tank cross
    section.(4) Mechanical agitators (motor-driven impellers) provide positive
    mixing action and are widely used where maintenance is possible.
    b. Obstacles to Effective Mixing
    Precise measurement requires that the composition and temperature
    of the solution to be measured be completely uniform before the solution
    is sampled. Unfortunately for this purpose, vessels containing solutions
    of SNM are often, for criticality prevention, made in the shape of
    slender cylinders, thin slabs, or thin annuli-shapes that are not
    suitable for effective mixing. Cylindrical vessels whose axes are
    horizontal or nearly so are especially poorly adapted to efficient
    mixing.(5) If a slab or cylinder does not have adequate storage
    capacity for a given volume of liquid and if several such slabs or
    cylinders are manifolded in parallel, the problem is further aggravated.
    When several such vessels are manifolded in parallel, uniformity of the
    contents of all of the vessels is difficult to achieve.
    Boron-containing raschig rings used as tank filters to prevent
    criticality are another impediment to effective mixing.
    ----------
    (*) Recommended procedures and equipment for taking liquid samples
    are contained in another regulatory guide (in preparation) in this
    series.
    ----------
C. REGULATORY POSITION
    This section provides guidance for vessel design to facilitate the
    accurate measurement of contained SNM.
1. The following guidelines are of general applicability to the
    design of vessels in which accountability measurements are to be made.
    a. The vessel design should take into account the
    elements of random and systematic error contributed by each component of
    the bulk measurement and sampling operations and should minimize the
    overall measurement error resulting therefrom.
    b. The vessel design should consider the means for
    standardizing and calibrating each measurement operation(*) and should
    provide for continuing quality assurance of the measurement system
    during plant operation.
    c. To arrive at the overall vessel design, the designer
    should integrate the bulk measurement and sampling operations with all
    process requirements and with any additional constraints imposed by
    considerations of safety and criticality prevention.
    d. Piping connected to measurement vessels should be
    designed to drain reproducibly and, if possible, to drain quickly.
    e. The design should require that only with the
    measurement vessel and its contents at rest should measurements be made.
2. The following guidelines apply to vessels whose bulk content
    is determined by weighing the vessel and its contents.
    a. The tank should be installed in such a manner that no
    variable extraneous loads are imposed as a result of mechanical or
    thermal stresses in attached piping. Systems having essentially zero
    deflection (null balance) should be considered as a means for reducing
    stress effects from connecting piping. Even with a null-balance system,
    the vessels should be protected from thermal and mechanical forces.
    Care should be taken to minimize rigidity of piping to weighing vessels,
    as by the use of properly supported flexible metal lines or
    small-diameter piping with reasonably long horizontal runs. To minimize
    thermal stresses, the design should provide for constant or reproducible
    temperatures in the connecting piping insofar as possible.
    b. Convolutions in flexible metal lines and any bends,
    incorporated to minimize piping rigidity, should be oriented so that the
    piping drains freely.
    ----------
    (*) See, when published, a regulatory guide (in preparation),
    "Measurement Control Program for Special Nuclear Material Accounting."
    See also ANSI Standards N15.18, "Mass Calibration Techniques for Nuclear
    Materials Control," and N15.19, "Volume Calibration Techniques for
    Nuclear Materials Control" (both in preparation).
    ----------
    c. Mechanical linkages between the weighing device and
    readout mechanism should be minimized, and measurements should be
    recorded by means of remote printout, particularly if the alternative is
    a series of levers to transfer the reading from the vessel to the
    measuring location. In any case, digital printout is recommended in
    order to avoid parallax errors, misreading, and recording blunders. If
    an integrated computerized inventory system is used, the weight should
    be recorded by the computer.
    d. Any mechanical linkages used in the load transmission
    system should be lubricated and protected from corrosion or the
    introduction of dust.
    e. If a heating-cooling jacket is needed on the solution
    tank, the piping should be designed so that either the jacket can be
    filled to a constant volume or it can be drained completely to eliminate
    errors caused by varying weights of fluid in the jacket.
    f. The weighing system should be designed so that initial
    calibration and periodic recalibration can be done in place. This
    requires that the system have the capability of adding precisely known
    weights to the vessel over the entire measurement range.
    g. Since vessels mounted on weighing systems are not
    supported against shocks or earthquakes, appropriate use should be made
    of staybars or limit stops.
3. The following guidelines apply to vessels whose bulk content
    is determined by measuring volume, temperature, and specific gravity.
    a. The vessel should be calibrated by liquid additions in
    such a manner that the total error of the bulk volume calibration at any
    liquid level is within the accuracy appropriate to its overall
    contribution to the limit of error of material unaccounted for.
    (Greater accuracy in calibration is normally required for vessels
    containing feed or product solutions than for vessels that contain waste
    solutions.) The calibration should include a sufficient number of
    liquid levels to bracket closely all discontinuities or fluctuations in
    cross-sectional area.
    b. The following features in the design of a vessel
    enhance the capability of obtaining overall measurement errors that are
    acceptable:
    (1) The vessel should be a right circular cylinder,
    with the principal axis oriented vertically. Vessels having a slab
    configuration should be avoided unless special provisions are made for
    ensuring uniformity of cross section with height.
    (2) The ratio of vessel height to horizontal
    cross-sectional area should be such that the product of the
    cross-sectional area and the smallest increment of liquid level that can
    be measured does not exceed the required volume sensitivity.
    (3) For cylindrical vessels fabricated of flat
    stock, care should be taken to ensure that the circumference is uniform
    with cylinder length. Anomalous circumference variations in cylindrical
    vessels should not exceed 0.05 percent between calibration points.
    (4) The vessel interior should be as free as
    possible of coils, baffles, and other objects that interfere with the
    linearity of the calibration.
    (5) When mixing baffles or other internal devices
    are essential to the design, such baffles or devices should have a
    uniform cross section throughout the range where measurements are
    normally made. Two or three such constant-area regions are permissible
    in a vessel, provided the transition areas are outside normal operating
    regions.
    c. Provision should be made for periodic in situ
    recalibration of vessels by the addition of precisely known quantities
    of liquid.(6) d. For vessels in which absolute quantities are measured,
    provision should be made for keeping the heel volume constant, and the
    heel volume should be kept small. To the extent feasible, the heel
    volume ordinarily should not exceed 1% of the nominal volume of liquid
    measured.
    e. Whether mixing is done by sparging or by an impeller,
    measurement of the liquid level should be done when the vessel contents
    are quiescent in order to get an accurate reading. During measurement,
    recirculating samplers should be turned off and the gas flow to the dip
    tubes should be the minimum practical.
    f. If pneumatic dip tubes are used for determining liquid
    level, care should be taken in selecting the diameter of the dip tubes,
    the length of gas supply lines, the shape of the open end (bubbler
    orifice), the fixing of the dip tubes in the tank, and the materials of
    construction; also, equalization of flow rates in any pair of dip
    tubes(7) should be provided.
    The dip tubes should be installed so that the opening of one
    of them is located above the highest level reached by the liquid and any
    foam formed on top of the liquid. The lower dip tube should be located
    as low as practicable.
    The lower tube should be rigid and should be supported to
    resist bending. The openings at the ends of the tubes and the pneumatic
    supply lines should be designed so that pressure drop between tube ends
    and the point where pressure is measured does not exceed about 3 pascals
    (0.01 in. H(2)O) for the flow rated used. Pneumatic supply lines should
    be leak-tight.
    To provide a means for measuring the specific gravity of the
    liquid, an additional dip tube may be installed with its open end in the
    liquid a fixed distance above the lower tube. This measurement can be
    used to confirm any specific gravity determinations on a sample of the
    liquid contents.
    The type of readout, manometer or transducer, should be
    evaluated for each dip-tube application. A manometer gives a more
    direct measurement. Also, a precision manometer read by a careful,
    trained operator gives better accuracy than can be obtained with
    transducers routinely associated with process instrumentation. However,
    a precision pressure transducer can give readings that are just as
    accurate-without the need for a trained operator. Furthermore, the
    transducer readings can be transmitted directly to a data collection
    system, thereby avoiding reading errors entirely. However, transducers
    should be calibrated frequently. For either type of instrument, the
    design should provide for in-place calibration.
    If an integrated computerized inventory system is used, the
    differential pressures and the equivalent solution bulk should be
    recorded by the computer.
    g. The weight factor should not be used without
    temperature correction unless it can be demonstrated that any change in
    specific gravity is exactly offset by a change in liquid depth. This
    condition will occur only if the vessel cross section is constant with
    changing temperature or if the changes encountered in vessel volume or
    temperature are small enough that thermal expansion of the vessel walls
    can be neglected.
    h. If liquid level is measured by means of a sight glass,
    care should be taken to ensure that the density of the liquid in the
    sight glass is the same as the density of the liquid within the vessel.
    The scale attached to the sight glass should be of noncorrosive material
    and should have a low coefficient of thermal expansion; provision should
    be made for its calibration. The designer should also ensure that the
    sight glass will measure all liquid levels to be encountered during its
    use. Protection should be provided against breakage of the sight glass.
    i. If the use of several thin slabs or long cylinders in
    parallel is necessitated by criticality considerations, provision should
    be made for ensuring the homogeneity of the contents of all of the
    vessels. Alternatively, provision should be made for ensuring that
    solution cannot leak from one vessel to another and that solution
    intended for a given vessel cannot leak (e.g., through a valve) to
    another vessel in the bank. Provision should also be made for measuring
    the volume of liquid in each vessel, and each vessel in the bank should
    be capable of being sampled independently.
4. The following guidelines apply to design considerations
    pertaining to mixing of vessel contents to ensure the obtaining of a
    representative sample.
    a. A single gas sparger may be used for vessels up to 3 m
    (10 ft) in diameter; larger vessels should have more than one sparger.
    b. Use of raschig rings in vessels in which
    accountability measurements are to be made should be avoided wherever
    possible. Other means of preventing criticality should be considered
    for such vessels. If raschig rings are used, precautions should be
    taken to ensure thorough mixing and completeness of transfer. Because
    raschig rings tend to compact during use, the designer should consider
    the need for more frequent calibration of vessels filled with raschig
    rings.
    c. If mixing is done by external recirculation and if the
    recirculating pump and piping are not dedicated to the measurement
    vessel, care should be exercised that the solution to be measured is
    completely returned to the measuring vessel and that none escapes
    elsewhere in the process. In any case, the returning solution should be
    distributed in the vessel volume-for example, by spraying.
D. IMPLEMENTATION
    This section provides information to applicants and licensees
    regarding the NRC staff's plans for using this regulatory guide.
    Except in those cases in which the applicant proposes an
    alternative method for complying with specified portions of the
    Commission's regulations, the method described herein will be used in
    the evaluation of submittals in connection with special nuclear material
    license, operating license, or construction permit applications docketed
    after June 1, 1975.
    If an applicant whose application for a special nuclear material
    license, an operating license, or a construction permit is docketed on
    or before June 1, 1975, wishes to use this regulatory guide in
    developing submittals for applications, the pertinent portions of the
    application will be evaluated on the basis of this guide.
    REFERENCES
1. D. M. Considine and S. D. Ross, "Handbook of Applied
    Instrumentation," pp. 5-41 to 5-54, McGraw-Hill, New York, 1964.
2. D. M Considine, ed., "Process Instruments and Controls Handbook,"
    pp. 7-8 to 7-23, McGraw-Hill, New York, 1957.
3. F. M. Groth and F. O. Cartan, "Evaluation of Instrumentation for
    Nuclear Fuels Reprocessing Plant Input Weight Measurements," USAEC
    Report ICP-1014, July 1972.
4. J. T. Long, "Engineering for Nuclear Fuel Reprocessing," pp.
    330-332, Gordon and Breach, New York, 1967.
5. J. E. Harrell, "Mixing and Sampling Enriched U-235 Fluids in
    Cylindrical Storage Containers," USAEC Report Y-1561, January 17,
    1967.
6. C. G. Rodden, ed., "Selected Measurement Methods for Plutonium and
    Uranium in the Nuclear Fuel Cycle," USAEC Report TID-7029, 2d
    edition, pp. 61-65, 1972.
7. J. T. Long, op. cit., pp. 733-736.
    53